378
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_rfanalyzer.fm
RFANalyzer
RFANalyzer:MANual:FREQuency
Function
Sets/queries the frequency that the mobile station is expected to transmit on. See
The units (HZ | KHZ | MHZ | GHZ) are optional, if no units are specified then units default to
HZ.
Setting the manual frequency changes the receiver control to manual. see
“RFANalyzer:CONTrol:AUTO” on page 372
Range
Range: 292.5 MHZ to 2700 MHZ
Resolution: .01 HZ
Query
Range: 292.5 MHZ to 2700 MHZ
Resolution: .01 HZ
*RST setting
896 MHZ
Programming Example
OUTPUT 714;”RFANALYZER:MANUAL:FREQUENCY 942.6MHZ” !Sets the expected frequency
!to 942.6 MHZ in manual
!receiver mode.