![Agilent Technologies 8960 Series 10 Reference Manual Download Page 490](http://html1.mh-extra.com/html/agilent-technologies/8960-series-10/8960-series-10_reference-manual_2868627490.webp)
490
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_SYSTem_FTRigger.fm
SYSTem:FTRigger
SYSTem:FTRigger
“Diagram Conventions” on page 213
SYSTem:FTRigger:BIT
SYSTem:FTRigger:STATe
Function
Selects/queries which bit, after zero, will be used for frame trigger pulse positioning. See
Frame Trigger Parameters” on page 515
Setting
Range: 0 to 1250
Resolution: 1
Query
Range: 0 to 1250
Resolution: 1
*RST Setting
zero
Programming Example
OUTPUT 714;”SYSTEM:FRTIGGER:BIT 14” !Would cause external frame trigger pulses
!to occur 14 bits after bit 0 of the
!selected timeslot.
Function
Sets/queries the frame trigger state. See
“Setting Frame Trigger Parameters” on page 515
Setting
0|OFF | 1|ON
Query
0|1
*RST Setting
0|off
Programming Example
OUTPUT 714;”SYSTEM:FRTIGGER:STATE ON” !Set frame trigger state to on.
SYSTem
:FTRigger
<sp>1|ON|0|OFF
?
<sp><num value>
?
returns 1|0
?
<sp><num value>
Summary of Contents for 8960 Series 10