SETup:ORFSpectrum
SETup:ORFSpectrum:SWITching:FREQuency[:OFFSet]
Function
Sets/queries the list of output RF spectrum due to switching frequency offsets. Each offset listed
in the command is turned on by default. If no frequency offsets (null list) are sent, the output RF
spectrum due to switching measurement will not be made. The units (GHZ|MHZ|KHZ|HZ) are
optional, if no units are specified than units default to HZ. See
Measurement Description” on page 75
Setting
Range: 0 to 8 comma-separated values ranging from
−
1.8 MHz to
−
10 Hz, and +10 Hz to +1.8
MHz
Resolution: 10 Hz
Query
Range: 0 to 8 comma-separated values ranging from
−
1.8 MHz to
−
10 Hz, and +10 Hz to +1.8
MHz
Resolution: 10 Hz
*RST Setting
Offset 1 = 400.0 kHz
Offset 2 = 600.0 kHz
Offsets 3 to 8 off
Programming Example
OUTPUT 714;"SETUP:ORFSPECTRUM:SWITCHING:FREQUENCY:OFFSET 400 KHZ, 700 KHZ"
!Turns on the first two ORFS due to switching measurement offsets and sets them
!to 400 kHz and 700 kHz offsets. All other offsets are in the off state.
OUTPUT 714;"SETUP:ORFSPECTRUM:SWITCHING:FREQUENCY:OFFSET 700 KHZ" !Turns on the
!first ORFS
!due to
!switching
!measurement
!offset and
!sets it to
!700 kHz
!offsets. All
!other offsets
!are in the
!off state.
OUTPUT 714;"SETUP:ORFSPECTRUM:SWITCHING:FREQUENCY:OFFSET" !Turns all of the ORFS
!due to switching
!measurements offsets
!to off.