Step 6a: Start Set Of Concurrent Measurements
Step 6a: Start Set Of Concurrent Measurements
Starting Measurements
The INITiate command is used to start measurements. Each individual measurement in a test application can
be started using the INITiate command. For starting measurements, the syntax of the INITiate command is as
follows:
Example 1. Command Syntax:
INITiate:<meas-mnemonic>[:ON]
The following table shows the measurements available in the Agilent Technologies E1960A GSM mobile test
application and their associated <meas-mnemonic> used in the INITiate command syntax.
Example 2. Programming Example:
OUTPUT Test_set;"INIT:TXP"
would start the transmitter power measurement.
Using Compound Commands to Start Multiple Measurements
More than one measurement can be started using a single INITiate command. For example:
OUTPUT Test_set;"INIT:TXP;PFER"
would start the transmit power measurement and the phase and frequency error measurement. These
measurements would then run concurrently.
Measurement Mnemonics Used In The INITiate Subsystem
Measurement
<meas-mnemonic>
Transmit Power
TXPower
Power vs Time
PVTime
Phase & Frequency Error
PFERror
Output RF Spectrum
ORFSpectrum
Bit Error
BERRor
Fast Bit Error
FBERror
Decoded Audio
DAUDio
Analog Audio
AAUDio
I/Q Tuning
IQTuning
Dynamic Power
DPOWer