SETup:CONTinuous
SETup:CONTinuous
February 14, 2000
“Diagram Conventions” on page 213
SETup[:ALL]:CONTinuous
Function
Sets trigger arm to OFF (single trigger) or ON (continuous trigger) for all measurements. See
“Trigger Arm (Single or Continuous) Description” on page 151
At power on and a (manual user) full preset the trigger arm is set to continuous. Partial preset
has no effect on the trigger arm state. See
“Preset Descriptions” on page 535
Remote full preset sets the trigger arm to single, this is the recommended trigger arm for any
remote measurements.
Trigger arm may be set and queried for each individual measurement.
Setting
Range
• Continuous trigger = ON
• Single trigger = OFF
*RST Setting
Single
Programming Example
OUTPUT 714;"SETUP:ALL:CONTINUOUS:OFF" !Sets trigger arm for all measurements
!to single.
OUTPUT 714;"SETUP:PVTIME:CONTINUOUS:OFF" !Sets trigger arm for power versus
!time measurements to single each
!measurment can be set individually.
OUTPUT 714;"SETUP:TXPOWER:CONTINUOUS?" !Queries trigger arm for TX power
!measurements. Trigger arm may be queried
!only one measurement at a time.
SETup
:OFF|:ON
[:ALL]