344
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_fetch_pvtime.fm
FETCh:PVTime
FETCh:PVTime:POWer[:ALL]:AVERage?
FETCh:PVTime:POWer[:ALL]:SDEViation?
Function
Queries the average power levels, from a number of multi-measurements, at each user-settable
time offset that is currently turned on. Results are relative to the power versus time carrier
power measurement.
The
command sets up the number of offsets that are turned on
and their time values. The
queries the number of offset points
that are turned on, indicating the number of values to expect when you send this command.
Query
Range:
−
100 dBc to +10 dBc and 9.91 E+37 (NAN)
Resolution: 0.01 dB
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
Function
Queries the standard deviation, from a number of multi-measurements, at each user-settable
time offset that is currently turned on.
The
command sets up the number of offsets that are turned on
and their time values. The
queries the number of offset points
that are turned on, indicating the number of values to expect when you send this command.
Query
Range: 0 dBc to +100 dBc and 9.91 E+37 (NAN)
Resolution: 0.001 dB
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX