SETup:AAUDio
SETup:AAUDio:EXPected:VOLTage[:PEAK]
SETup:AAUDio:FILTer[:SFRequency]
SETup:AAUDio:FILTer:FREQuency
Function
Sets/queries the maximum expected peak voltage (clipping level) of the Analog Audio signal to be
measured. The units (V|MV|UV) are optional, if no units are specified then units default to V.
see
Setting
Range: 10 mV to 20 V peak / Resolution: 0.1 mV
Query
Range: 10 mV to 20 V peak / Resolution: 0.1 mV
*RST setting
20V
Programming Example
OUTPUT 714;”SETUP:AAUDIO: EXPECTED:VOLTAGE 5 V” !Sets the clipping level of
!Analog Audio measurements
!to +5 volts.
Function
Sets/queries the state to on and the center frequency for the 100 Hz bandpass filter applied to
Analog Audio measurements. Units (KHZ|HZ) are optional, if no units are specified then units
default to Hz. see
“Analog Audio Measurement Description” on page 44
Setting
Range: 200 Hz to 8.0 kHz / Resolution: 1 Hz
Query
Range: 200 Hz to 8.0 kHz / Resolution: 1 Hz
*RST setting
1000 Hz
Programming Example
OUTPUT 714;”SETUP:AAUDIO:FILTER :SFREQUENCY 1000” !This is a complex command that
!sets the aaudio filter state
!to on and sets the bandpass
!filter frequency to 1 kHz.
Function
Sets/queries the center frequency for the 100 Hz bandpass filter applied to Analog Audio
measurements. Units (KHZ|HZ) are optional, if no units are specified then units default to Hz.
see
Setting
Range: 200 Hz to 8.0 kHz / Resolution: 1 Hz
Query
Range: 200 Hz to 8.0 kHz / Resolution: 1 Hz
*RST setting
1 kHZ
Programming Example
OUTPUT 714;”SETUP:AAUDIO:FREQUENCY 217HZ” !Set aaudio bandpass filter to 217 hz.