CALL:POWer
CALL[:CELL]:POWer:AMPLItude
CALL[:CELL]:POWer:STATe
Function
Sets/queries the Cell Power of the test set, this is the same for BCH and TCH.
The suffix dBm is optional.
The Cell Power field is affected when there is an amplitude offset, see
Setting
range: -10 dBm to -127 dBm
resolution: .01 dBm
Query
range: -10 dBm to -127 dBm
resolution: .01dBm
*RST Setting
-85 dBm
Programming Example
OUTPUT 714;”CALL:CELL:POWER:AMPLITUDE -50dBm” !Set the cell power from test
!set to -50dBm.
Function
Sets/queries the RF Power state.
Setting
range: 0 | OFF | 1 | ON
Query
range: 0 | OFF | 1 | ON
*RST Setting
ON
Programming Example
OUTPUT 714;”CALL:CELL:POWER:AMPLITUDE -50dBm” !Set the cell power from test
!set to -50dBm.