![Agilent Technologies 8960 Series 10 Reference Manual Download Page 425](http://html1.mh-extra.com/html/agilent-technologies/8960-series-10/8960-series-10_reference-manual_2868627425.webp)
425
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_pferror.fm
SETup:PFERror
SETup:PFERror:TRIGer:SOURce
SETup:PFERror:TRIGger:QUALifier
Function
Sets/queries the trigger source for phase/frequency measurements. See
Error Measurement Description” on page 82
Range
AUTO|PROTocol|RISE|IMMediate
See
“Triggering of Measurements” on page 149
Query
AUTO | PROT | RISE | IMM
*RST Setting
AUTO
Programming Example
OUTPUT 714;"SETUP:PFERROR:TRIGGER:SOURCE AUTO" !Sets trigger source to AUTO.
Function
Selects/queries the trigger qualifier for phase/frequency measurements. See
Setting
Range: 0|OFF | 1|ON
Query
0|1
*RST Setting
1|on
Programming Example
OUTPUT 714;"SETUP:PFERROR:TRIGGER:QUALIFIER OFF" !Sets trigger qualifier state
!to off.
Summary of Contents for 8960 Series 10