358
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_initiate.fm
INITiate
INITiate:DONE:FLAG<measurement mnemonic>
INITiate:ON?
Function
Specifies which measurements are considered, (included or excluded) when the DONE? query is
sent.
If a measurements trigger arm, see
“Trigger Arm (Single or Continuous) Description” on page
, has been left in continuous mode, the done flag for that measurement will toggle between
DONE and WAIT, see
. The INITiate:DONE? query will probably
not be able to catch the measurement at the instant it is done, therefore the measurement will
never appear to be done. If a measurement trigger arm must be left in continuous mode the user
should (exclude) it, using this command, from the INITiate:DONE? query results.
Once the INITiate:DONE:FLAG has been set to EXCLude for a measurement, the user must
send the INCLude command for that measurement in order to query that measurement with, the
INITiate:DONE? query. The test set will not reset any (excluded measurement) to be an (included
measurement) with any form of preset, see
“Preset Descriptions” on page 535
Setting
Range
• INCLude
:AAUDio| :BERRor | :DAUDio | :DPOWer | :FBERror | :IQTuning | :ORFSpectrum |
:PFERror | :PVTime |:TXPower
• EXCLude
:AAUDio| :BERRor | :DAUDio | :DPOWer | :FBERror | :IQTuning | :ORFSpectrum |
:PFERror | :PVTime |:TXPower
Related Topics
.
Programming Example
OUTPUT 714;”INITIATE:DONE:FLAG:AAUDIO EXCLUDE” !excludes AAUDIO measurements
!from contributing the
!INIATIATE:DONE? query.
Function
Queries the names of the measurements (none, one, or more than one) that are ON in a comma
separated list of measurement mnemonics. See
“INITiate:<measurement mnemonic>[:ON]” on
.
Query
Range: AAUD | BERR | DAUD | DPOW | FBER | IQT | ORFS | PFER | PVT | TXP | NONE