+100 to +199 Core Device-Specific Error
+113 Internal error; <VI NAME> forced inactive
Indicates that a VI is inactivated when not executed.
<VI NAME> could be : “IntVmVI”, “GSMFixedVI”,
“MiscVI”, or “GSMSacchMriVI”
+114 Internal error; <VI NAME> not responding
Indicates that a VI has not been instantiated or the
state is not available.
+120 Warning; Receiver overrange due to present setting
of amplitude offset (SYST:CORR:GAIN)
Indicates the combination of Expected Power and
Amplitude Offset are out of range for the test set.
+121 Warning; Receiver underrange due to present
setting of amplitude offset (SYST:CORR:GAIN)
Indicates the combination of Expected Power and
Amplitude Offset are out of range for the output level
of the MS.
+122 Warning; Reference out of lock
Indicates the test set’s internal reference is out of lock.
+123 Warning; Duplicate RF IN/OUT Amplitude Offset
Frequency entry. First frequency entry in RF IN/OUT
Amplitude Offset table will be used.
Indicates that an amplitude offset value for that
frequency has already been entered. The test set will
use the amplitude offset value entered first.
+130 Configuration error; Unable to switch to indicated
Test Application
The test application selected is not available for this
test set.