99
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\prog_rach_proc.fm
Programming a RACH Measurement
Programming a RACH Measurement
This section provides an example of how to make a power versus time measurement on a RACH. The same
principles as used in this example can also be used for transmit power and phase and frequency error
measurements.
Overview of Measurement Procedure
1. Ensure that the mobile is switched off.
2. Set the test set’s measurement receiver to the frequency the RACH will arrive on. The simplest way to do
this is to set the manual channel (that is, the expected ARFCN) to the ARFCN of the BCH. Alternatively
you could set the expected frequency to the uplink frequency of the BCH ARFCN.
3. Set triggering to single.
4. Set trigger mode to Auto.
Once the RACH measurement is completed, in order to make further measurements on the TCH, ensure you
reset the receiver control to Auto.
NOTE
Only one measurement at a time can be made on the RACH even if two measurements are
initiated.