47
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\ts_aaud.fm
AAUDio Troubleshooting
AAUDio Troubleshooting
Possible Setup Issues
During remote operation of the analog audio measurement the user should configure the trigger arm to single,
see
“SETup:AAUDio:CONTinuous” on page 381
.
Failure to set trigger arm to single may result in the measurement never giving a result. When trigger arm is
continuous the measurement rearms itself and starts another measurement cycle, during remote operation
the fetch query may not be synchronized to the measurement cycle, see
“Measurement States” on page 150
.
The analog audio measurement results are rms values, the Expected Peak Audio Amplitude is a peak value.
Interpreting Integrity Indicator values
See
“Integrity Indicator” on page 125
.
If over range (5) is returned then the input level is greater than 3dB above the Expected Peak Audio
Amplitude value or the maximum input level of 20 volts peak.
If under range (6) is returned then the input level is greater than 20dB below the Expected Peak Audio
Amplitude value maximum value.
If the signal has both over range and under range conditions only the over range (5) will be indicated.