S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\prog_prog_guide_step6c.fm
Step 6c: Obtain a Set of Measurement Results
Overview
The FETCh subsystem is used to query measurement results. The measurement results from each
measurement in a test application can be queried using the FETCh subsystem. The general hierarchy of the
FETCh command structure is as follows:
FETCh:<meas-mnemonic>:<result format>?
The following table shows the measurements available in the Agilent Technologies E1960A GSM mobile test
application and their associated <meas-mnemonic> used in the FETCh command syntax.
The command syntax used to obtain the various measurement result formats (<result format>) for each
measurement can be found in the test set’s FETCh? subsystem’s GPIB command syntax reference
information.
Example 1. Command Syntax:
FETCh:<meas-mnemonic>:<result format>?
Example 2. Programming Example:
OUTPUT Test_set;"FETCh:TXP:POW:MIN?"
would return the minimum value from the set of samples taken during the transmit power measurement
(when averaging is turned on and number of samples taken >1).
Measurement Mnemonics Used In The FETCh Subsystem
Measurement
<meas-mnemonic>
Transmit Power
TXPower
Power vs Time
PVTime
Phase & Frequency Error
PFERror
Output RF Spectrum
ORFSpectrum
Bit Error
BERRor
Fast Bit Error
FBERror
Decoded Audio
DAUDio
Analog Audio
AAUDio
I/Q Tuning
IQTuning
Dynamic Power
DPOWer
Summary of Contents for 8960 Series 10