S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_fetch_orfspectrum.fm
FETCh:ORFSpectrum
FETCh:ORFSpectrum:SWITChing:FREQuency[:OFFSet]:SDEViation?
FETCh:ORFSpectrum:SWITChing[:ALL]:SDEViation?
Function
Queries the ORFS due to switching measurement, allowing frequency offset values to be
appended to the command. Returns ORFS due to switching (standard deviation) measurements
at the frequencies listed, in the order they are listed.
Frequencies must have a one-to-one correspondence to ORFS due to switching frequency offsets
that are currently turned on. Frequencies must be separated by commas. (See
“SETup:ORFSpectrum:SWITching:FREQuency[:OFFSet]”
for the command that turns on
frequency offsets.)
Each value is (optionally) followed by: HZ|KHZ|MHZ|GHZ . The default units are HZ (hertz).
Query
Range: 0 dB to +150 dB and 9.91 E+37
Resolution: 0.001 dB
Programming Example
OUTPUT 714;”FETCH:ORFSPECTRUM:SWITCHING:FREQUENCY:OFFSET:STDEVIATION? 200 KHZ, 400 KHZ”
!Returns the ORFS due to switching (standard deviation) measurement
!results at the 200 kHz and 400 kHz offsets only, assuming these
!offsets are turned on.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
Function
Queries output RF spectrum due to switching (standard deviation) measurement results at all
frequency offsets currently turned on (there can be up to eight).
The
“SETup:ORFSpectrum:SWITching:FREQuency[:OFFSet]”
command sets up the number of
output RF spectrum due to switching offsets that are turned on and their frequency values. The
“SETup:ORFSpectrum:SWITching:FREQuency:POINts?”
queries the number of output RF
spectrum due to switching points that are turned on, indicating the number of output RF
spectrum due to switching values to expect when you FETCh results.
Query
Range:
−
100 dBm to +100 dBm and 9.91 E+37
Resolution: 0.01 dB
Programming Example
OUTPUT 714;”FETCH:ORFSPECTRUM:SWITCHING:ALL:SDEVIATION?” !Returns the ORFS due
!to switching
!(standard
!deviation)
!measurement
!results at all
!frequency offsets
!currently turned on.
XXXXXXXXX