Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
2-7
Measurement Mode
Sweep Measurement Mode
Subordinate Sweep Measurement
For subordinate sweep measurement, you set up a secondary sweep source (VAR2)
in addition to a primary sweep source (VAR1). After primary sweep is completed,
the output of secondary sweep source is incremented or decremented by the
specified step value, then the primary sweep source is swept again.
Figure 2-2
Subordinate Sweep Measurement
To set up the subordinate sweep measurement, select the following on CHANNELS:
CHANNEL DEFINITION page:
•
VAR1 function for desired primary sweep source (SMU or VSU).
•
VAR2 function for desired secondary sweep source (SMU or VSU).
Subordinate sweep measurement has the following restriction:
•
For the secondary sweep source, only
single
sweep mode and
linear
staircase
mode are available.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...