8-40
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
If You Have A Problem
If Errors Occur
STRESS: STRESS FORCE screen
Status on STRESS: STRESS FORCE screen is displayed in following format:
STATUS:
A C
Where,
A
and
C
mean as follows:
A
hardware status error. If multiple errors occur, numbers are
added and displayed as a hexadecimal number.
•
2 : Oscillation.
•
4 : Some channel has reached compliance limit.
C
PGU status
•
1 : PGU average current exceeds 100 mA.