Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
4-5
Making a Measurement
Connecting DUT
Connections for High Current Measurements
When you force or measure a large current, you may want to use a
Kelvin (4-wire)
connection
to eliminate the residual resistance effects of test leads and contacts. For
example, you can use the following connections as Kelvin connections on the test
fixture. The Kelvin connection is available for the 4156B’s HRSMU and
41501A/B’s HPSMU.
Examples: Kelvin Connection
To cancel the effects of the residual resistance, test leads must be connected as close
as possible to the DUT.
SMU(±200V= 1A Max)
Sense
Force
G u a r d
G u a r d
1
Sense
Force
G u a r d
G u a r d
2
Sense
Force
G u a r d
G u a r d
3 (Selected)
Gate
Source
Drain
Substrate