3-40
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Trigger Function
Setup and restrictions
•
You
cannot
perform trigger outputs together with trigger inputs. You
must
select
either trigger output or trigger input.
•
To use a trigger function, you must enable the trigger function and select either
TRIG OUT or TRIG IN in the TRIGGER SETUP table on the MEASURE:
OUTPUT SEQUENCE screen. Then the trigger inputs or outputs are performed
automatically after you start a measurement by selecting a measurement
front-panel key (Single, Repeat, or Append).
•
The
trigger output
function is
not
available for sampling measurements.
•
When you perform knob sweep measurements, the trigger function is not
available.
•
For the electrical specifications of trigger signals, refer to
User's Guide General
Information
.