1-22
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Units
Pulse Generator Unit (PGU)
Leading-edge and trailing-edge transition time
The leading-edge and trailing-edge transition times have five setting ranges as
shown in Table 1-16.
Table 1-16
Ranges and Resolutions of Leading and Trailing Transition Time
•
restrictions
•
leading-edge transition time
£
pulse width
´
0.8.
•
trailing-edge transition time
£
(
pulse period
-
pulse width
)
´
0.8.
•
Leading and trailing-edge transition times for a PGU must be in the same
range.
Output impedance
You can select 50
W
or Low impedance.
Trigger output
PGUs output trigger signal to synchronize with external pulse generators. If an
41501A/B has PGUs, the 41501A/B has a trigger output terminal. For details of
trigger functions, refer to “Trigger Function” in Chapter 3.
Range
Leading and Trailing
Transition Time
Resolution
1
100 ns to 1000 ns
1 ns
2
0.50
m
s to 10.00
m
s
10 ns
3
5.0
m
s to 100.0
m
s
100 ns
4
50
m
s to 1000
m
s
1
m
s
5
0.5 ms to 10.00 ms
10
m
s
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...