Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
3-59
Measurement Functions
Compliance
Table 3-7
V Compliance Resolution
Table 3-8
I Compliance Setting Range
HPSMU
1 nA to 10 mA
0 to 200 V
100 mA (0
£
| I |
£
50 mA)
0 to 200 V
100 mA (50 mA < | I |
£
115 mA)
0 to 100 V
1 A (0
£
| I |
£
50 mA)
0 to 200 V
1 A (50 mA < | I |
£
125 mA)
0 to 100 V
1 A (125 mA < | I |
£
500 mA)
0 to 40 V
1 A (500 mA < | I |
£
1 A)
0 to 20 V
Unit V
Compliance
Resolution
HRSMU
0 V
£
| V |
£
2 V
100
m
V
MPSMU
2 V < | V |
£
20 V
1 mV
HPSMU
20 V < | V |
£
40 V
2 mV
40 V < | V |
£
100 V
5 mV
100 V < | V |
£
200 V
10 mV
Unit
Output Range
I Compliance Setting Range
HRSMU
2 V
100 fA to 100 mA
20 V
100 fA to 100 mA
40 V
100 fA to 50 mA
100 V
100 fA to 20 mA
MPSMU
2 V
1 pA to 100 mA
20 V
1 pA to 100 mA
40 V
1 pA to 50 mA
100 V
1 pA to 20 mA
Unit Output
Range
V Compliance
Setting Range
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...