7-42
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Data Variable and Analysis Function
Analysis Function
Overlay Display Function
You can overlay a measurement curve (that was previously saved into one of the
four internal memories) onto the curve that is presently displayed on the
GRAPH/LIST: GRAPHICS screen. This is useful for comparing measurement
results.
Overlay Display Information
You can use following information of overlaid curve instead of present information:
•
Axis information
•
Cursor and marker position (x, y1, y2)
•
Line x interrupt, y1 interrupt and gradient, y2 interrupt and gradient
•
List of the data variables
Adjusting axes
You can use the axis scaling of overlaid plane instead of present scaling.
Automatic Analysis Function
This function can automatically draw up to two lines and position a marker on the
plotting area of the graph screen. You set up this function on the DISPLAY:
ANALYSIS SETUP screen. This function is performed automatically when:
•
measurement finishes.
•
AUTO ANALYSIS secondary softkey is pressed.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...