3-8
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Stress Force Function
NOTE
Pulse Waveform when Stress Stops
When you set the duration mode or press the Stop front-panel key, be aware that
stress force may stop during the pulse peak output as shown in the following figure:
Sequence for returning to 0 V (stress force state to the idle state)
When the state changes from the stress state to the idle state, the outputs of the
channels are returned to 0 V in opposite order that forcing occurred.
Delay time of PGUs
When PGUs are set to
VPULSE
(ac source), you can set a delay time as follows:
•
If PGU is set to
SYNC
, the PGU waits the delay time (after the
stress
start trigger
is received), then starts to force ac stress.
•
If PGU is set to
NSYNC
, the PGU waits the delay time (after stress
force
state
starts), then starts pulse output.
Example
Figure 3-1 shows an example of output sequence when forcing stress.
Figure 3-1 assumes the output sequence is set on the MEASURE: OUTPUT
SEQUENCE screen as follows.
1. PGU1
2. SMU1
3. PGU2
4. SMU2
5. SMU3
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...