6-66
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Screen Organization
GRAPH/LIST Screen Group
SPREADSHEET softkey
Select SPREAD SHEET primary softkey to display ASCII SAVE window. The
following entry fields appear:
Also, the following softkeys appear:
•
Select EXECUTE softkey to store result data to diskette file or network disk.
•
Select EXIT softkey to exit the ASCII SAVE window.
•
Select FILE CATALOG secondary softkey to list the names of all files that are
on diskette or network disk. You can select a file name from the list.
ASCII SAVE function automatically adds
TXT
extension to specified file name.
•
NAME
Enter the name of file (without extension) to which you want to save the result
data.
•
OUTPUT DATA
Enter numbers to specify range of data you want to save. These numbers
correspond to
NO.
column of LIST screen.
•
right field: upper limit
•
left field: lower limit
Select ALL secondary softkey to specify all result data.
•
UNIT
Specify whether to include units (for example,
V
or
ms
).
•
ON secondary softkey to include units.
•
OFF secondary softkey to not include units.
FUNCTION:ASCII SAVE
OUTPUT DATA (INDEX NO)
NAME
UNIT
DELIMITER
STRING MARK
<-->
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...