Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
2-23
Measurement Mode
Sampling Measurement Mode
•
SEQUENTIAL mode
Source units starts output in the order defined in the OUTPUT SEQUENCE
table of the MEASURE: OUTPUT SEQUENCE screen. Time Origin is when
the last source reaches the specified output value. See figure below.
If there is pulse bias sources (PGUs), they start to force pulse base value in the
order shown above, and start to force pulse bias at the Time Origin.