2-32
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Mode
Sampling Measurement Mode
Rule to determine @TIME:
@TIME value is determined by the following rule. Data measured at @TIME=
Tlog
are stored in memory.
Tlog
³
Ttarget
|
Tlog
-
Ttarget
| < |
Ttarget
-
Tprev
|
where,
Tlog
Data stored in @TIME. Actual measurement point.
Ttarget
Target value of @TIME. The values can plot data on the logarithmic
X-axis in the same interval completely.
Tprev
Tlog
-
sampling interval
. Actual measurement point.
A= |
Ttarget
-
Tprev
|
B = |
Tlog
-
Ttarget
|
TIME
Tprev
Tlog
Ttarget
Ttarget < Tlog
A > B
Sampling
Interval
Discard
Stored in @TIME
Measurement
B
A
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...