Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
2-11
Measurement Mode
Sweep Measurement Mode
Pulse Sweep Measurement
For a sweep measurement, a sweep or constant source SMU can be a pulse source.
But
only one
SMU can be a pulse source. Figure 2-4 shows the relationship between
pulse source and other sources.
Figure 2-4
Pulse Source and Other Sources
For the pulse sweep measurement, the delay time of the primary sweep source is
ignored, and each step of the primary sweep source is synchronized with output of
the SMU pulse source. Measurements are made during the pulse output.
The pulse output of PGU is not synchronized with any other source.