Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
3-47
Measurement Functions
SMU/PG Selector Control
Setup and Switching Conditions
Setup of the SMU/PG SELECTOR field and switching conditions are explained
below:
CH1 (or CH3 for second selector):
CH2 (or CH4 for second selector):
NOTE
SW1 and SW2 are mechanical relays, and SW3 is a semiconductor switch. Leakage
current and stray capacitance of SW1 and SW2 are less than for SW3. However, the
switching speed of SW3 is faster and life is longer than SW1 and SW2.
If you need to switch PGU many times, use PGU OPEN, not OPEN to disconnect
PGU.
Restrictions using Selector
•
41501A/B Expander equipped with PGU must be connected to 4155B/4156B.
•
Kelvin connection is not available.
Setup
SW1 SW2 SW3
Description
SMU
ON
OFF
OFF
Connects SMU.
PGU
OFF
ON
ON
Connects PGU.
PGU OPEN
OFF
ON
OFF
Open. Disconnected.
OPEN
OFF
OFF
OFF
Open. Disconnected.
Setup
SW1 SW2
Description
SMU
ON
OFF
Connects SMU.
PGU
OFF
ON
Connects PGU.
OPEN
OFF
OFF
Open. Disconnected.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...