6-10
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Screen Organization
CHANNELS Screen Group
•
If both PGUs are set to VPULSE, the STBY settings of both PGUs must be
same.
FCTN.
This field defines an output function for SMUs, VSUs, PGUs, and GNDU. When
the pointer is located in this column, allowable output functions appear in the
secondary softkey area. You select a softkey to set an output function.
•
SMU or VSU: you can set VAR1, VAR1', VAR2, or CONST.
•
PGU or GNDU: you can set CONST.
Restrictions:
•
In FCTN column, you
cannot
set multiple VAR1, VAR1', or VAR2. For
example, you
cannot
set VAR1 for 2 units.
•
If VAR1' is set, you must set VAR1 also.
•
If VAR2 is set, you must set VAR1 also.
•
The output modes of VAR1 and VAR1' must be same. That is, the MODE setting
for both must be set to a voltage mode, or both must be set to a current mode.
For example, you can set VAR1 to V and VAR1' to VPULSE.
•
You
cannot
set VAR1, VAR1', or VAR2 for sampling measurement. You can set
CONST only.
STBY.
STBY field specifies which channels output source values in the standby state.
•
If STBY is set to ON, the unit forces a specified output value when in the
standby state.
•
If STBY is blank, the unit outputs 0 V in the standby state (same as when in idle
state).
See “Types of Operation State” in Chapter 3 for more information on the standby
state.
Restrictions:
•
If both PGUs are set to VPULSE, the STBY setting of both PGUs must be the
same.
•
For STBY=ON channel, SERIES RESISTANCE setting must be 0 ohm.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...