4-22
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Making a Measurement
Sweep Measurements
Pulse Parameters
The relation between the PERIOD, WIDTH, and BASE values are as shown in the
following figures.
SMU outputs the pulses as shown in figure (a) or figure (b).
•
Figure (a)
When the function (FCTN) is set to VAR1, VAR2, or VAR1’.
The pulse peak values are the sweep output values calculated from the sweep
start, stop, step values, and so on.
•
Figure (b)
When the function (FCTN) is set to CONST.
The pulse peak value is the output value of the constant output source.