Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
4-13
Making a Measurement
Sweep Measurements
VNAME and INAME
You can use VNAME and INAME names in user function definitions or for analysis
on the GRAPHICS/LIST screens. These names must be 6 or less alphanumeric
characters. First character must be alphabet character.
To disable a unit
Select DELETE ROW softkey. The settings for the unit are deleted.
Example
The following settings show an example for measuring an n-p-n transistor's I-V
characteristics. SMU1 is connected to base, SMU2 is connected to collector, and
SMU3 is connected to emitter. SMU1 is set to current source (I mode) and
secondary sweep (VAR2) function. SMU2 is set to voltage source (V mode) and
primary sweep (VAR1) function. SMU3 is set to COMMON.