Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
3-35
Measurement Functions
Output Sequence
Sequential Mode
Default output sequence in the sequential mode is shown below. In the default
settings, output channels start the output in this order, and stop the output in the
opposite order.
1. SMU1
2. SMU2
3. SMU3
4. SMU4
5. VSU1
6. VSU2
7. PGU1
8. PGU2
Starting Outputs
In the idle state, output switches of
all
units are on, and the units output 0 V. When
moving to the measurement, stress force, or standby state, the units operate as
shown below:
1. Disabled units keep 0 V output, and turn the output switch off.
where disabled unit means the unit you select the DELETE ROW softkey in the
CHANNELS table on the CHANNELS: CHANNEL DEFINITION screen.
2. Output channels start the output in the order specified in the OUTPUT
SEQUENCE table.
Stopping Outputs
When returning to the idle state, the units operate as shown below:
1. Output channels stop the output in the opposite order of the OUTPUT
SEQUENCE table.
2. Disabled units keep 0 V output, and turn the output switch on.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...