2-6
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Mode
Sweep Measurement Mode
start
Start value of sweep. For logarithmic sweep,
start
must not be
zero. Allowable range of
start
depends on output range of
sweep source. For output range of each measurement channel,
refer to Chapter 1.
stop
Stop value of single sweep or turning back value of double
sweep. For logarithmic sweep,
stop
must have same polarity as
start
, and must not be zero. Allowable range of
stop
depends on
output range of sweep source. For output range of each
measurement channel, refer to Chapter 1.
step
•
For linear sweep,
step
is step increment of sweep. Number
of sweep steps is calculated from
start
,
stop
, and
step
.
Calculated number of steps must be in range: 2 to 1001.
•
For logarithmic sweep,
step
is invalid. Number of sweep
steps is calculated from
start
,
stop
, and number of steps in
one decade, which is specified by
log
parameter. Calculated
number of steps must be in range: 2 to 1001.
compliance
Compliance value of sweep source. This parameter applies to
SMU only. Allowable range of
compliance
depends on the
compliance range of sweep source. For the compliance range of
each measurement channel, refer to Chapter 1.
power compliance
(Optional) Power compliance value of sweep source. This
parameter applies to SMU only. Allowable range depends on
power compliance range of sweep source. For details, refer to
Chapter 3.
hold time
Time required for DUT to settle after forcing start value.
Allowable range is 0 to 655.35s. Resolution: 10 ms.
delay time
Time required for DUT to settle after stepping the output.
Allowable range: 0 to 65.535 s. Resolution: 100
m
s
Refer to “CHANNELS: CHANNEL DEFINITION screen” and “MEASURE:
MEASURE SETUP screen” in Chapter 6 for setting up these parameters.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...