4-60
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Making a Measurement
Stress Force
To Use Selector
1. Press Stress key in the PAGE CONTROL key group.
2. Select CHANNEL DEF primary softkey.
3. In the MEASURE field of the desired channel in the SMU/PG SELECTOR area,
select:
•
SMU secondary softkey to connect SMU to DUT during measurement state.
•
PGU secondary softkey to connect PGU to DUT during measurement state.
•
OPEN secondary softkey to disconnect SMU, PGU, and DUT during
measurement state.
•
PGU OPEN secondary softkey to disconnect PGU from DUT by
semiconductor switch during measurement state.
4. In the STRESS field of the desired channel in the SMU/PG SELECTOR area,
select:
•
SMU secondary softkey to connect SMU to DUT during stress force state.
•
PGU secondary softkey to connect PGU to DUT during stress force state.
•
OPEN secondary softkey to disconnect SMU, PGU, and DUT during stress
force state.
•
PGU OPEN secondary softkey to disconnect PGU from DUT by
semiconductor switch during stress force state.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...