1-12
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Units
Source/Monitor Unit (SMU)
Figure 1-5
HPSMU Output and Measurement Ranges
Table 1-9
HPSMU Output Voltage Ranges and Resolutions
Range Output
Value
Output
Resolution
Current
Compliance
Range
2 V
0
£ ½
V
½ £
2 V
100
m
V
±1000 mA
20 V
0
£ ½
V
½ £
20 V
1 mV
±1000 mA
40 V
0
£ ½
V
½ £
40 V
2 mV
±500 mA
100 V
0
£ ½
V
½ £
100 V
5 mV
±125 mA
200 V
0
£ ½
V
½ £
200 V
10 mV
±50 mA