4601 I-V Meter Operation Manual
4.2.10 MENU Key (Parameter Setting)
4-9
3)
Load @ PWR On
Selects the setting parameter condition at power ON.
Saved @ PWR Off Start up using the parameters saved when the power was
last turned off.
User-0 Saved
Start up using the parameters saved in the non-volatile
memory, User-0.
H)
INTERFACE
Selects and sets the interface.
1)
Interface Bus
Selects either the GPIB interface or the USB interface.
GPIB interface:
2)
GPIB Address
Sets the GPIB address from 0 to 30.
GPIB interface:
2)
USB Id
Sets the USB ID from 1 to 127.
3)
Header
Sets the header of the talker format to ON or OFF.
4)
Output Format
Selects the output format between ASCII and BINARY.
5)
Output Vm Data
Sets whether the Vm data (ADC1 measured value) is output or not.
On:
Output
Off:
No output
6)
Output Im Data
Sets whether the Im data (ADC2 measured value) is output or not.
On:
Output
Off:
No output
7)
Output ADC3 Dt
Sets whether the ADC3 measured value is output or not.
On:
Output
Off:
No output
I)
MAINTENANCE
Sets the maintenance and inspection-related parameters.
1)
Self Test
Executes the self test.
For more information, refer to Section 5.7, "Self Test."
2)
Disp/Key Test
Executes a test for the display and the panel keys.
For more information, refer to Section 5.7, "Self Test."
3)
Calib. Mode
Selects the calibration mode.
On:
Sets the status in which source values and measured values are
calibrated.
For more information, refer to Chapter 8., "CALIBRATION."
Off:
Sets the normal measurement status.
4)
Counter Select
Selects the relay counter.
5)
Relay Counter
Displays the relay ON/OFF count stored in the selected relay counter.
6)
Reset Counter
Clears the selected relay counter.
7)
Serial Number
Displays the serial number.