4601 I-V Meter Operation Manual
6.6.3 Remote Command List
6-28
sweep
Memory sweep
source data
N
[adr]
P
Sweep memory data setting starts from N command and com-
pletes at P command.
N<adr>, SOV<data1>, SOV<data2>, •••, P
adr:
Memory address (0 to 3999)
data1:
Voltage source value at adr
(0)
data2:
Voltage source value at adr+1
(0)
[NOTE]
The range cannot be specified.
*6
N?
[adr]
Response:
N<adr>, SOV
<data>, P
adr: <dddd>
data: <d.ddddE
d>
*1
NP?
Query of the source memory setting status
Response:
0•••source memory setting complete
1•••source memory setting in progress
RSAV
Saves the memory sweep data.
RLOD
Loads the memory sweep data.
RCLR
Initializes the memory sweep data.
(Data saved in memory is not initialized.)
Bias value
SB
[data]
data:
Bias value
(0)
SB?
Response: SB
<d.ddddE
d>
*1
RTB
(Return To Bias)
RB0
OFF (Stays at the final value when sweep stops.)
RB1
ON (Returns to the bias value when sweep stops.)
RB?
Response:
RB0 or RB1
Step count per
trigger
SPN
[cnt]
cnt:
Step count per trigger (1 to 2000)
(1)
SPN?
Response:
SPN<cnt> cnt: <dddd>
Reverse mode
SV0
OFF
SV1
ON
SV?
Response:
SV0 or SV1
Sweep stop
SWSP
Stops the sweep.
Trigger
*TRG
Sweep start trigger
Measurement trigger
Mea-
sure-
ment
Function
(ADC3
measurement)
F0
Measurement OFF
F3
DC current measurement Ir
(Reference cell Isc)*7
F4
External voltage measurement Em
(EXT.DCV)
F5
Temperature measurement Tc1
Ch1 thermocouple T
F6
Temperature measurement Pt1or Ad1 Ch1 PT100 or Ch1
AD590
F8
Temperature measurement Tc2
Ch2 thermocouple T
F9
Temperature measurement Pt2 or Ad2 Ch2 PT100 or Ch2
AD590
F?
Response:
F0, F3 to F6, F8, F9
Measurement
range
(ADC3
measurement)
R1
R2
R3
R?
Response:
R1 to R3
Pt/Ad selection
ATS0
Pt1,Pt2
ATS1
Ad1,Ad2
ATS?
Response:
ATS0 or ATS1
*1:
The response decimal point is different depending on the set value.
For the source value, limit value, and time parameter set up range, refer to Chapter 9, "SPECIFICATIONS.".
*6:
Not initialized by the RINI or *RST command.
*7:
Ir is set when switching to the sweep mode. It is not changed to another function in the sweep mode.
Item
Command
Description
Default
Operation
During
calibration
Power
ON
Factory
shipment
DC
OPR/SUS
Sweep
OPR/SUS
DCI (ISC)
DCV (EXT.DCV)
3 mA
30 mV
30 mA
300 mV
300 mA
3 V