4601 I-V Meter Operation Manual
6.6.3 Remote Command List
6-27
Sweep
2-slope linear
sweep
SLW
[
fd,
sd,
ld, st1, st2]
fd:
First value
(-1 mV)
sd:
Second value
(0 V)
ld:
Last value
(2 mV)
st1:
Step 1 count (1 to 1999) st1+st2
1999
(10)
st2:
Step 2 count (1 to 1999) st1+st2
1999
(10)
If all the settings are omitted, set the sweep type only.
However it is not allowed to omit each value separately.
SLW?
Response:
SLW
<fd>,
<md>,
<ld>, <st1>, <st2> *1
fd,md,ld,: <d.ddddE
d>
st1,st2: <dddd>
3-slope linear
sweep
SLR
[
fd,
sd,
td,
ld, st1, st2,
st3]
fd:
First value
(-1 mV)
sd:
Second value
(0 V)
td:
Third value
(1 mV)
ld:
Last value
(2 mV)
st1:
Step 1 count st1+st2+st3
1999
(10)
st2:
Step 2 count st1+st2+st3
1999
(10)
st3:
Step 3 count st1+st2+st3
1999
(10)
If all the settings are omitted, set the sweep type only.
However it is not allowed to omit each value separately.
SLR?
Response:
SLR
<fd>,
<sd>,
<td>,
<ld>, <st1>, <st2>,<St3>*1
fd, sd, td, ld: <d.ddddE
d>
st1, st2, st3: <dddd>
Fixed level
sweep
SFX
[
lvl, cnt]
lvl:
Level source value
(0 V)
cnt:
Sampling count (1 to 2000)
(1)
If all the settings are omitted, set the sweep type only.
However it is not allowed to omit each value separately.
SFX?
Response: SFX
<lvl>, <cnt>
lvl: <d.ddddE
d>
*1
cnt: <dddd>
Memory sweep
SMD
[st, sp]
st:
Start address (0 to 1999)
(0)
*3
sp:
Stop address (0 to 1999)
(0)
If all the settings are omitted, set the sweep type only.
However it is not allowed to omit each value separately.
SMD?
Response:
SMD st, sp
st,sp: <dddd>
Sweep type
SX?
Returns the current sweep type.
Response:
Linear sweep: Same as the response of SLN?
Fixed level sweep: Same as the response of
SFX?
Memory sweep: Same as the response of
SRD?
2-slope linear sweep: Same as the response of
SLW?
3-slope linear sweep: Same as the response of
SLR?
*1:
The response decimal point is different depending on the set value.
For the source value, limit value, and time parameter set up range, refer to Chapter 9, "SPECIFICATIONS.".
*3:
When sweep stops in the Operate status, the values can be changed only between the start and stop address that was set while it is in the Standby or Suspend status.
Item
Command
Description
Default
Operation
During
calibration
Power
ON
Factory
shipment
DC
OPR/SUS
Sweep
OPR/SUS