4601 I-V Meter Operation Manual
5.2.5 Source and Measurement Timings
5-14
2. Relation between setting restrictions and source modes
The following table shows the setting restrictions to be checked according to the source mode.
3. Source delay time and measurement delay time
The setting resolutions of the source delay time and the measurement delay time are determined by
the period time resolution. The source delay time and the measurement delay time operate with the
rounded off values.
(The setting values are displayed on the screen.)
The following table shows the minimum setting value for each resolution.
Setting restriction
Source mode/Sampling
DC
SWP/AUTO
SWP/HOLD
[(Td+300
s) < Tp]
-
-
[10 ms
Tp]
-
-
[200
s
Tp]
-
-
[Tds
Td]
-
[(Td+Tit+Tk) < Tp]
-
Period time setting range
Resolution
0.050 ms
to
60.000 ms
1
s
60.01 ms to
600.00 ms
10
s
600.1 ms
to 6000.0 ms
100
s
Time parameter
Period time resolution
1
s
10
s
100
s
Source delay time (Tds)
10
s
20
s
200
s
Measurement delay time (Td)
20
s
20
s
200
s