4601 I-V Meter Operation Manual
5.3.6 Zero Measurement
5-24
5.3.6
Zero Measurement
Zero measurement is a function that measures analog common voltage and cancels the offset errors
of the AD converters automatically.
Execute zero measurement once after warm-up.
1. Types of zero measurement
Types of AZ data (excluding the integration time)
AZ data to be used (excluding the integration time)
2. Timing of zero measurement
Zero measurement is performed when being specified by the ZM command or menu operation
(Refer to E) MEASURE in Section 4.2.10).
Zero measurement is performed internally when the integration time or measurement function
range is changed or spot measurement is performed.
No.
ADC
Function
Range
Zero code
1
ADC1
Vm
Voltage measurement
5 V-300 V
AZ10
2
ADC2
Im
Current measurement
300
A-10 A AZ20
3
ADC3
Ir
Current measurement
3 mA-300 mA AZ30
4
Em
Voltage measurement
30 mV
AZ31
5
300 mV
AZ32
6
3 V
AZ33
7
Tc1,Tc2 Temperature measurement
54 mV
AZ34
No.
ADC
Function
Range
Zero code
1
ADC1
Vm
Voltage measurement
5 V-300 V
AZ10
2
ADC2
Im
Current measurement
300
A-10 A
AZ20
3
ADC3
Ir
Current measurement
3 mA-300 mA AZ30
4
Em
Voltage measurement
30 mV
AZ31
5
300 mV
AZ32
6
3 V
AZ33
7
Tc1
Temperature measurement
55 mV
AZ34
8
Tc2
Temperature measurement
55 mV
AZ34
9
Ad1
Temperature measurement
3 V
AZ33
10
Ad2
Temperature measurement
3 V
AZ33
1
2
3
1
2
3
1
2