4601 I-V Meter Operation Manual
3.1 Power ON
3-2
3.1
Power ON
Turn On the 4601 according to the following procedure:
Ensure that the voltage of the used power supply matches the power voltage display on the rear panel.
Then, connect the supplied power cable
CAUTION:
To prevent damage to the 4601, do not apply a voltage or frequency that exceeds the specified range.
Press the
POWER
switch to turn on the power.
After the panel and all the indicators go ON, the self-test is automatically executed.
When a problem occurs during the self-test, an error message is displayed.
For more information on the error contents, refer to Section A.2, “Error Message List.”
When the self-test terminates correctly, the screen is displayed as follows for about three seconds.
The start up process is complete, and the DC source mode screen appears.
CAUTION:
1.
The 4601 starts up with the initial values that are preset before shipment when it is turned on for the first time.
When the setting parameter condition at Power ON is changed, the screen display may be different from the
above figure.
For more information, refer to Section 3.6.3, “Auto Parameter Loading at Power ON.”
When turning on the output for the first time, pay attention to the source value setting and the load connection.
2.
If a problem occurs during the self-test, an error message will appear on the screen.
Refer to Section A.2, “Error Message List.” to solve the problem.