4601 I-V Meter Operation Manual
9. SPECIFICATIONS
9-1
9.
SPECIFICATIONS
Every overall accuracy is satisfied at a temperature of 23
5
C and a relative humidity not exceeding 85%,
and with the integration time of 1 PLC or longer, including the calibration accuracy, the 1-day stability, the
temperature coefficient, and the linearity.
However, the stated temperature coefficients are reference values.
9.1
Voltage Source/Voltage Measurement/Current Measurement Terminal
OUTPUT/SENSE Terminal
Voltage source/measurement range:
Current limit/current measurement range:
With the integration time of 1 PLC or longer
*1:
The polarities of current limits and current measurement are represented as “+” for source and as “-”
for sink.
The above ranges are applied to the current limit setting ranges and the current measurement ranges;
however the output range is limited to +0.1 A and -10.2 A.
The measurement resolution with the integration time of 500
s or less is as follows:
Overall accuracy
Voltage source/voltage measurement
Range
Source range
Setting
resolution
Measurement range
Measurement
resolution
5 V
-1.0000 V to
+5.0000 V
100
V
-1.00999 V to
5.00999 V
10
V
50 V
-1.000 V to
+50.000 V
1 mV
-1.0999 V
to +50.0999 V
100
V
300 V
-1.00 V to +300.00 V
10 mV
-1.999 V
to +300.999 V
1 mV
Range
Limit setting range *1
Limit setting
resolution
Measurement range *1
Measurement
resolution
300
A
3
A
to 320
A
100 nA
0 to
320.999
A
1 nA
3 mA
30
A to
3.2 mA
1
A
0 to
3.20999 mA
10 nA
30 mA
0.3 mA to
32 mA
10
A
0 to
32.0999 mA
100 nA
300 mA
3 mA to
320 mA
100
0 to
320.999 mA
1
A
3 A
30 mA to
3.2 A
1 mA
0 to
3.20999 A
10
A
10 A
0.1 A to
10.2 A
10 mA
0 to
10.2999 A
100
A
Integration time
5
s
10
s
25
s
50
s
100
s
250
s
500
s
Measurement
resolution (digits)
17
9
4
2
2
2
1
Range
Voltage source
Voltage measurement
(% of s V)
(% of r V)
5 V
0.025+1 mV
0.025+500
V
50 V
0.025+10 mV
0.025+2 mV
300 V
0.025+100 mV
0.025+20 mV