4601 I-V Meter Operation Manual
4.2.10 MENU Key (Parameter Setting)
4-7
Fixed level sweep:
1)
Bias Value
Sets the bias value.
2)
Level Value
Sets the level value.
3)
Sample Count
Sets the sampling count.
Memory sweep:
1)
Bias Value
Sets the bias value.
2)
Sweep Address
Sets the start address and the stop address.
3)
Memory Data Set
Sets the source value at each memory address.
4)
Mem. Save/Clear
Saves or clears the memory data.
D)
TIME
Sets the time parameters.
1)
Integ. Time
Sets the integration time IT.
2)
Source Delay
Sets the source delay time Tds.
3)
Measure Delay
Sets the measurement delay time Td.
4)
Period
Sets the sampling interval in the DC source mode or the step time Tp in the
sweep source mode.
5)
Hold Time
Sets the hold time Th in the sweep source mode.
6)
Trigger Delay
Sets the delay time Tdr between the external trigger (TRIGGER IN signal) and
the sweep start or between the external trigger and the next step value output.
E)
MEASURE
Sets the measurement-related parameters.
1)
Memory Recall
Recalls and displays data from the measurement data memory.
2)
Memory Store
Selects whether measured data is stored to the measurement data memory or
not (ON or OFF) in the DC source mode.
Memory store is always set to ON in the sweep source mode.
3)
Memory Clear
Clears data in the measurement data memory.
4)
Measure Zero
Measures necessary internal zero points to cancel zero point errors in mea-
sured values. Zero measurement should be executed after warm up.
5)
Measure Switch
Switches measurement ON or OFF.
6)
Pt/Ad Select
Selects the sensor to be connected to the temperature measurement connector.
7)
RTD
Sets the resistance bulb standard.
Pt100
Pt100 (compliant with JIS C1604-1997/IEC 60751)
JPt100
JPt100 (compliant with JIS C1604-1981)
8)
Temp. Unit
Sets the temperature unit.
C,
F or K