4601 I-V Meter Operation Manual
6.6.3 Remote Command List
6-29
Mea-
sure-
ment
Integration time
IT0
5
s
IT1
10
s
IT2
25
s
IT3
50
s
IT4
100
s
IT5
250
s
IT6
500
s
IT7
1 ms
IT8
2.5 ms
IT9
5 ms
IT10
10 ms
IT11
1 PLC
IT12
2 PLC
IT13
100 ms
IT14
200 ms
IT?
Response:
IT0 to IT14
Spot measure-
ment
integration time
SIT7
1 ms
SIT8
2.5 ms
SIT9
5 ms
SIT10
10 ms
SIT11
1 PLC
SIT12
2 PLC
SIT13
100 ms
SIT14
200 ms
SIT?
Response:
SIT7 to SIT14
Spot measure-
ment
execution
SPM?1
Vm
Executes the specified function measurement
and returns the measured value.
SPM?2
Im
rng: 1 to 3 (can be omitted) For more
information, refer to the R command of the
measurement range setting
SPM?3 [,rng]
Ir
SPM?1: ADC1 Vm measurement
SPM?4 [,rng]
Em
SPM?2: ADC2 Im measurement
SPM?5
Tc1
SPM?3 to SPM?10: ADC3 specifying
function measurement
SPM?6
Pt1 or Ad1
SPM?8
Tc2
The response is output in the format
described in Section “6.5.4 ASCII Output
Format.”
SPM?9
Pt2 or Ad2
However, the header is always output in the
ASCII format.
NOTE:
Not executable in the DC source mode.
Measurement
buffer
memory
ST0
Store
OFF
*8
ST1
ON
ST?
Response:
ST0 or ST1
RL
Initializes the stored data.
RN
n[,adr]
n:
0... Releases recall execution status.
*5
1... Sets recall execution status.
adr:
Recall data number (0 to 3999)
(The data number is not changed if this
setting is omitted.)
Reading out the recall data by using the talker
function after the recall execution status is set
performs the following operation.
• Increments the recall data number after
the data output.
• If data does not exist in the specified
number, the output
becomes <EE +8.30>.
• Reading out does not erase the data in
memory.
*5:
Not initialized by the RINI command.
*8:
It is set to OFF when switching to the DC mode, and set to ON when switching to the sweep mode.
It cannot be set to OFF in the sweep mode.
Item
Command
Description
Default
Operation
During
calibration
Power
ON
Factory
shipment
DC
OPR/SUS
Sweep
OPR/SUS