4601 I-V Meter Operation Manual
8.3 Calibration Points and Tolerance Ranges
8-8
2. Voltage measurement terminal
3. Reference cell measurement terminal
External voltage measurement calibration (Em)
Range
Calibration point
Tolerance
specification [digit]
30 mV
00.0000 mV
d
+30.0000 mV
d
300 mV
000.000 mV
d
+300.000 mV
d
3 V
0.00000 V
d
+3.00000 V
d
Reference cell current measurement calibration (Ir)
Range
Calibration point
Tolerance
specification [digit]
3 mA
0.00000 mA
d
+3.00000 mA
d
30 mA
00.0000 mA
d
+30.0000 mA
d
300 mA
000.000 mA
5d
+300.000 mA
d