4601 I-V Meter Operation Manual
5.6.4 Serial Sweep
5-44
The output voltage is the total of No.1 output voltage (HI side) and No2. output voltage (LO side).
Set the current limit values under the following conditions:
No.1 current limit value < No.2 current limit value
When setting the limit values to 10 A, set as follows:
No.1 limit value = 10 A
No.2 limit value = 10.2 A
When the PV output voltage exceeds 300 V, set the output status to Operate according to the following
procedure.
If this procedure is not followed, an overvoltage alarm may occur.
<Operate procedure>
No.2 bias value
PV output voltage - 300 V
No.1 bias value = 0 V
Set No.2 to Operate.
Set No.1 to Operate.
Start sweep.
As for series connection, select LoZ when using Suspend.
Figure 5-14 shows serial sweep connection.