7.7.5 Battery type
Your insulin pump has the option of using 1.5 volt AA Alkaline batteries or
rechargeable AA NiMH batteries. If you change from one battery type to
the other, you must change the battery type accordingly in the SETUP
MENU STANDARD.
See the section 4.3 “Battery” for further information on recommended
batteries.
Press
d
to move to the SETUP MENU
STANDARD.
Select with
f
.
Press
d
to move to the BATTERY TYPE screen.
The current battery type is displayed.
Press
a
or
s
to select the desired battery
type (ALKALINE or NiMH [rechargeable] battery).
Save and exit with
f
.
Warning
If you do not set the correct battery type on your insulin
pump, the alert A2: BATTERY LOW may not occur in
time to provide you sufficient warning to replace your
battery.
112