![background image](http://html.mh-extra.com/html/accu-chek/spirit/spirit_user-manual_456003096.webp)
Note
When an Extended bolus, a MultiWave bolus and/or a tempo-
rary basal rate (TBR) is/are active and you put your insulin
pump into
STOP
, an alert A6: TBR CANCELLED and/or an
alert A8: BOLUS CANCELLED occur(s).
Press
f
twice for each appearing alert to confirm and
turn off.
See the sections 10.1.6 “Alert A6: TBR CANCELLED”
and 10.1.8 “Alert A8: BOLUS CANCELLED” for further
information.
Turn off the STOP-Warning
Press and hold
s
or
a
until you hear a melody. The
STOP
-Warning is
turned off.
Note
The volume of the
STOP
-Warning is independent of the
programmed beep volume. It always occurs at the maximum
level.
The
STOP
-Warning will be reactivated the next time your pump is
changed from
RUN
to
STOP
or with the next battery change.
STOP-Warning
The
STOP
-Warning is turned on each time your insulin
pump is put from
RUN
into
STOP
or when a battery is
inserted. To inform you that the insulin delivery is inter-
rupted, your insulin pump reminds you every minute
with a long beep and a vibration. Turning off the
STOP
-Warning suppresses this warning function.
94