Note
If your insulin pump was in
RUN
before you removed
the battery, an error E8: POWER INTERRUPT occurs.
Press twice
f
to confirm and to turn off the error.
Check the time and date in the
STOP
screen and correct,
if necessary.
Put your insulin pump into
RUN
, if necessary.
5.2 Start-Up procedure
Your insulin pump performs a start-up procedure (internal tests) when
you insert a battery and when you change the cartridge. When the battery
is changed the entire start-up procedure will be performed. When only
changing the cartridge, the start-up procedure will begin at the SELF-TEST
step.
Caution
When you remove the battery your insulin pump
remembers the settings you made and saved. If your
Accu-Chek Spirit insulin pump remains without a battery
for more than one hour, check the time and date upon
reinsertion of a battery.
Using AA batteries other than those supplied or
recommended by Roche Diagnostics may significantly
reduce the battery usage time. Batteries other than
those recommended may leak and corrode the battery
contacts within your insulin pump. Therefore, the use of
batteries not supplied or recommended by Roche
Diagnostics may void the warranty. Do not insert old or
used batteries into your insulin pump as this may cause
a failure at the insulin pump start-up.
48
Prepare the pump for use
Summary of Contents for Spirit
Page 1: ...INSULIN PUMP User Guide ...
Page 10: ...8 ...
Page 14: ...12 ...
Page 26: ...24 ...
Page 36: ...34 ...
Page 76: ...74 ...
Page 84: ...82 ...
Page 122: ...120 ...
Page 146: ...144 ...
Page 186: ...184 ...
Page 212: ...210 ...
Page 218: ...216 ...
Page 225: ......