10.2 Errors
10.2.1 Error E1: CARTRIDGE EMPTY
If the cartridge is empty, an error E1: CARTRIDGE EMPTY occurs. Change
the cartridge immediately.
Press
f
twice to turn off and confirm the error.
Disconnect or remove the infusion set from your
infusion site.
Change the cartridge and prime the new
infusion set.
Reconnect the infusion set (tubing) at your
infusion site or insert the new infusion set.
When finished, put your insulin pump into
RUN
.
Warning
After the occurrence of an error, your insulin pump
will be in
STOP
and the insulin delivery is interrupted.
In order to maintain insulin delivery, you must act
immediately according to the instructions given for each
error code.
152
Alerts and errors